Snježana Hrabrić Vlah [*] [1] Štefica Dvornik [1] Dino Grdović [1]

Show more about author

Introduction

 
Gem® PremierTM 4000 (Inc. Lexington, MA, USA) is a new analytical system designed for point-of-care testing (POCT), an analyzer that has appeared on the market lately. The term POCT designates biochemical testing out of a clinical laboratory, in surgical wards, intensive care units, emergency departments, etc. (1). POCT provides rapid findings and effective triage of patients (2,3).
The aim of this study was to compare a new Gem® PremierTM 4000 analyzer with those reported for reference POCT analyzers used until then, before its installation in a decentralized location (in the operating room). The use of a POCT analyzer is important to make fast and reliable clinical decisions while the patient is subjected to operation treatment (regarding transfusion, acid-base management, ventilation and glucose management). In addition, POCT analyzers eliminate time delay due to the transport of samples to the central laboratory and reduce the number of blood samples to measure parameters such as pH, pCO2, pO2, electrolytes, metabolites and hemoglobin/hematocrit (tHb/HCT) (4).
Special attention was given to technology of tHb/HCT concentration measurement on POCT analyzers. There are two types of technology for tHb/HCT measurement. Some POCT analyzers measure HCT using a technology called conductivity and calculate tHb (Gem® PremierTM 3000). An instrument like Ciba Corning 865 measures tHb using the technology called co-oximetry and calculates HCT. Gem® PremierTM 4000 instrument uses co-oximetry technology, and we wanted to find out which method is more accurate and more reliable for the measurement of tHb concentration. For clinicians, concentrations of tHb are significant for decision about blood transfusion.
 

Materials and methods

 
In this study, the analytical performance of the Gem® PremierTM 4000 was compared with the Ciba Corning 865 (Bayer Corporation, USA), Gem® PremierTM 3000 (Inc. Lexington, MA 02421, USA) and Dimension RxL (Dade Behring Inc, Newark, DE, USA) analyzers during the period of 30 days in the Rijeka Clinical Hospital Center, Rijeka, Croatia. These analyzers, except for Dimension RxL, were placed in the operating room and laboratory staff used them in their work.
Table 1 shows analytical methods of the Gem® PremierTM 4000 analyzer and comparable methods of other reference analyzers. pH, pCO2 and electrolytes were measured by potentiometry, pO2, glucose and lactate by amperometry. tHb concentrations were measured by co-oximetry and HCT was calculated (HCT% = 3.0 x tHb).
 
Table 1. Analytes and methods on Gem® PremierTM 4000 analyzer vs. comparative methods of other examined analyzers
 
19-2_Hrabric_Vlah_S._et_Al._Table_1
 
A total of 200 whole arterial blood samples were analyzed on the Gem® PremierTM 4000 analyzer and on Ciba Corning 865 and Gem® PremierTM 3000 (analytes: pH, pCO2, pO2, K, Na, glucose, tHb, HCT), as reference analyzers. In addition, 130 of those blood samples were analyzed for lactate and glucose. In the case of glucose, plasma samples (Li-heparin) were analyzed on the Dimension RxL as a reference analyzer. The blood samples that were used in the study were collected from adult patients for whom determination of pH, pCO2, pO2, electrolytes, metabolites and tHb/HCT was requested during operations so that no additional blood sampling was required. Whole blood samples were collected in Li-heparin syringes (BD Preset TM, Plymouth, U.K.) and analyzed immediately upon collection on all analyzers simultaneously. After having been injected into the Gem® PremierTM 4000, the bubble generated by the aspiration process was eliminated and the sample was injected into another analyzer. As for determinations of glucose, heparinized blood samples were centrifuged (10 min/3,500rpm) and plasma was analyzed on Dimension RxL analyzer.
Within-run imprecision was assessed by multiple (10 times) concentration determinations of pH, pCO2, pO2, electrolytes, metabolites, tHb and HCT in two different patient whole blood samples of low and high concentrations, respectively.
During the 10 days, between-day imprecision was assessed by determination of the same analyte concentration in two commercial control samples of two concentration levels (Rapid QC Complete Level 1 and Level 2, Bayer HealthCare LLC, MA,USA) once a day.

Statistical analysis

Statistical analysis was performed with MedCalc® (MedCalc 9.3.3.0, Mariakerke, Belgium). In order to determine imprecision, the arithmetical mean (X), standard deviation (SD) and coefficient of variation (CV %) were calculated. Correlation was expressed by Pearson’s correlation coefficient (r) with calculation (95%) of confidence interval (CI) for each of the investigated analytes. Passing and Bablok regression analysis (5) of results was used for the calculation.
Our acceptance criteria were: CV < 5% for imprecision, correlation coefficient r > 0.95 for assessment of the level of correlation of results.
 

Results

 
For pH, pCO2, pO2, electrolytes, metabolites, tHb and HCT, within-run imprecision was determined on the basis of results of 10 successive measurements at two concentration levels of different samples (levels I and II) (Table 2). Results of the within-run imprecision study were satisfactory for all analytes (CV ≤ 4.89%), except for pO2 (CV = 14.66%). Between-run imprecision was assessed by measuring analytes in two commercial controls for 10 successive days (Table 2). Between-run imprecision was satisfactory for all parameters (CV ≤ 4.41 %), with exception of lactate (Level I: CV = 8.0 %, Level II: CV = 6.59 %).
 
Table 2. Results of within-run and between-run imprecisions
 
19-2_Hrabric_Vlah_S._et_Al._Table_2
 
Our study of the Gem® PremierTM 4000 analyzer was conducted by comparison with reference (Dimension RxL, Ciba Corning 865 and Gem® PremierTM 3000) analyzers. The results of comparison studies of the Gem® PremierTM 4000 analyzer and reference analyzers are shown in Table 3, where Pearson’s correlation coefficients (r) are calculated. Correlation coefficients showed good correlation for most parameters (0.9458-0.9948), except for sodium (Gem® PremierTM 4000 vs. Ciba Corning 865; r = 0.8635). Correlation coefficients of tHb were excellent, r = 0.9923 (Gem® PremierTM 4000 vs. Ciba Corning 865) and r = 0.9759 (Gem®PremierTM 4000 vs. Gem® PremierTM 3000).
 
Table 3. Pearson’s correlation coefficients
 
19-2_Hrabric_Vlah_S._et_Al._Table_3
 
Results of compared determination of analyte concentrations in samples processed by Passing Bablok regression analyses are presented in Table 4. Results of the regression analysis showed good compatibility between Gem® PremierTM 4000 and reference analyzers, except for sodium (Gem® PremierTM 4000 vs. Ciba Corning 865; y = 34.7500 + 0.7500x) and tHb (Gem® PremierTM4000 vs. Gem® PremierTM 3000; y = 23.6923 + 0.8308x).
 
Table 4. Results of Passing Bablok regression analysis
 
19-2_Hrabric_Vlah_S._et_Al._Table_4
 

Discussion

 
On the basis of the results presented in Table 2, within-run and between-run imprecisions were satisfactory for all parameters (CV ≤ 4.89 %), except for pO2 where within-run imprecision was slightly higher (CV = 14.66%). Better results for between-run imprecision for pO2 than for within-run imprecision are expected due to multiple opening of syringe during sampling in within-run imprecision testing. The present criteria of between-run imprecision were not acceptable for lactate (Level I: CV = 8.0 %, Level II: CV = 6.59 %). Pearson´s correlation coefficients showed good correlation for most parameters (0.9458-0.9948), except for sodium (Gem® PremierTM 4000 vs. Ciba Corning 865; r = 0.8635). Results of the regression analysis showed good compatibility between Gem® PremierTM 4000 and reference analyzers, except for sodium (Gem® PremierTM 4000 vs. Ciba Corning 865; y = 34.7500 + 0.7500x). Ciba Corning 865 is the older type of POCT analyzer and we supposed that the electrode for the sodium was almost exhausted. This may be the reason for some unacceptable results for sodium.
Special attention was given to results of comparison of methods for determining the concentration of tHb between the new analyzer and reference analyzers, because tHb values are important to make decisions regarding blood transfusion during operation. There are two types of technology for tHb/HCT measurement (4). Gem® PremierTM 3000 analyzer measures HCT using a technology called conductivity and calculates tHb (tHb = 0.31 x HCT). Other similar instruments (Gem® PremierTM 4000 and Ciba Corning 865) measure tHb using the technology called co-oximetry and calculate HCT (HCT = 0.30 x tHb and HCT = 0.2941 x HCT). Correlation coefficient for tHb was excellent (r = 0.9923, Gem® PremierTM 4000 vs. Ciba Corning 865), as well as the result of the regression analysis (y = 1.0000 + 1.0000x, Gem® PremierTM 4000 vs. Ciba Corning 865), because both of them measure concentrations of tHb by co-oximetry. However, correlation coefficient for tHb was good (r = 0.9759, Gem® PremierTM 4000 vs. Gem® PremierTM 3000), but the result of the regression analysis was unacceptable (y = 23.6923 + 0.8308x, Gem® PremierTM4000 vs. Gem® PremierTM 3000), when tHb was calculated. Even if the correlation factors are within permitted limits, two measurements can still show discrepancy due to different regression analysis. The most probable reason for that is the difference between methods for determining concentrations of tHb between two analyzers (Gem® PremierTM 4000 and Gem® PremierTM3000). In determining tHb concentration, Gem Premier 4000 is completely comparable with Ciba Corning 865 analyzer, both analyzers using the same co-oximetry method (6).
The results of analytical performance of the Gem® PremierTM 4000 analyzer show that it is acceptable for daily application in the POCT (7,8), and the analyzer can be installed in decentralized locations.
 

Acknowledgment

 
We are grateful to Ventus-Medical d.o.o., Zagreb, for the provision of instruments and reagents for the analytical evaluation of the Gem® PremierTM 4000 analyzer.

Notes

Potential conflict of interest
None declared

References

1.   Di Serio F, Antonelli G, Trerotoli P, Tampoia M, Matarese M, Pansini N. Appropriateness of poin-of-care testing (POCT) in an emergency department. Clin Chem Acta 2003;333:185-9.
 2.   Di Serio F, Amodio G, Varraso L, Campaniello M, Coluccia P, Trerotoli P, et al. Integration between point-of-care cardiac markers in an emergency/cardiology department and central laboratory: methodological and preliminary clinical evaluation. Clin Chem Lab Med 2005;43(2):202-9.
 3.   Drenck N. Point of care testing in critical care medicine: the clinician´s view. Clin Chem Acta 2001;307:3-7.
 4.   Mayes GJ, Browne J. Point of care hematocrit and hemoglobin in cardiac surgery. Perfusin 2007;22(3):179-83.
 5.   Passing H, Bablok W. A new biometrical procedure for testing the equality of measurements from two different analytical methods. Application of linear regression procedures for method comparison studies in clinical chemistry. Part I. J Clin Chem Clin Biochem 1983;21:709-20.
 6.   Steinfelder-Visscher J, Weerwind PW, Teerenstra S, Brouwer MH. Reliability of point-of-care hematocrit, blood gas, electrolyte, lavtate and glucose measurement during cardiopulmonary bypass. Perfusio 2006;21(1):33-7.
 7.   Beneteau-Burnat B, Pernet P, Pilon A, Latour D, Goujon S, Feuillu A, Vaubourdolle M. Evaluation of the Gem® PremierTM 4000: a compact blood gas CO-Oximeter and electrolyte analyzer for poin-of-care and laboratory testing. Clin Chem Lab Med 2008; 46(2):271-9.
 8.   Vukelić N, Papić Futač D, Topić E. Analytical properties of the Gem Premier 3000 analyzer evaluated. Biochemia Medica 2007;17(2):231-41.