Introduction
Comparability of laboratory test results depends on standardization of all phases of laboratory testing, including pre-analytical, analytical and post-analytical phases. Pre-analytical and analytical phases of laboratory testing aim to generate an accurate test result, while the post-analytical phase - when the clinician receives the test results, interprets them, and uses them to make diagnostic and therapeutic decisions - aims to reduce errors or bias associated with the hand-off from laboratory to clinician. The most frequent errors in the post-analytical phase are erroneous validation of analytical data, failure to report test results to appropriate parties, excessively long turnaround time (TAT), mistakes in data entry, manual transcription errors and failure or delay in reporting critical values (1).
Despite the obvious importance of the post-analytical phase to overall laboratory performance, many providers of external quality assessment (EQA) schemes do not take into account the post-analytical phase (2). In 2009, the Croatian Chamber of Medical Biochemists (CCMB) and Croatian Society of Medical Biochemistry and Laboratory Medicine (CSMBLM) assessed the state of pre- and post-analytical procedures in medical laboratories across the country (3). The results indicated urgent, substantial need for improvement. Therefore, a nationwide EQA scheme covering the post-analytical phase was implemented in 2014, administered by the Croatian Centre for Quality Assessment in Laboratory Medicine (CROQALM) within the CSMBLM. The EQA scheme is implemented modularly three times per year. The CCMB made participation in the scheme mandatory for all medical laboratories in Croatia in 2013 (4). In the second EQA exercise of 2014, pilot modules on pre- and post-analytical phases were introduced; in all three EQA exercises of 2015, Module 11 dealing with the post-analytical phase was performed.
The present study was undertaken to evaluate to what extent the recently introduced nationwide EQA scheme for the post-analytical phase of laboratory testing has influenced laboratory practice in Croatia. Since laboratories showed substantial variation in post-analytical practices before the scheme (3), we felt it necessary to evaluate the success of the EQA scheme at this early stage in order to identify the more important issues and implementation gaps and thereby help regulators and laboratory directors focus their energies more efficiently in the coming years. In a separate publication, we will assess pre-analytical procedures using an EQA module for the pre-analytical phase developed by the CSMBLM.
Module 11 is an educational module about the post-analytical phase of laboratory testing, and it contains an optional questionnaire that presents medical laboratories with routine post-analytical scenarios where standardized practices exist under the Croatian EQA scheme or where clear rules are lacking (‘grey areas’). The present study retrospectively analysed laboratory responses to this questionnaire in 2014-2015 in order to (a) gain on-the-ground insights into current laboratory practices in Croatia and (b) identify the most urgent areas for improving the standardization of the post-analytical phase in Croatian laboratories.
Materials and methods
Study design
This retrospective, longitudinal study involved analysis of the responses of Croatian medical laboratories to a questionnaire distributed during four national EQA exercises conducted in September 2014 and in May, September and November 2015. During this study period, 194 medical laboratories were registered in the Croatian health care system, comprising 125 (64%) medical laboratories from primary health care facilities (including private medical practices and private laboratories) and 69 (36%) medical laboratories from secondary and tertiary health care centres (clinical hospital centres, clinical hospitals, general hospitals, national hospitals, and special hospitals). Although all medical laboratories were obliged to participate in the national EQA scheme, their responses on the questionnaire were voluntary. Laboratories were told that their responses would not affect their overall assessment from CROQALM. No fees or compensation were involved in completing the questionnaire.
Data from all responding laboratories were used in the present study; no exclusion criteria were applied. When they filled out the questionnaire, laboratories gave consent for the data to be stored and used by CROQALM for group-level analyses. Members of CROQALM signed statements that they would safeguard the confidentiality of EQA data.
Questionnaire
Questionnaires have been proposed as an effective method for assessing the post-analytical phase during EQA exercises (2). The Croatian EQA questionnaire was designed by CROQALM, approved by CSMBLM and CCMB, and distributed to all registered medical laboratories in the country. The responses were analysed by the EQA provider (CROQALM), and one of the authors (JLK) annotated the results in her capacity as EQA/CROQALM Module 11 coordinator. All these steps, from design to final analysis, were conducted via Web interface using inlab2*QALM software, specifically designed in 2011 for quality evaluation of medical laboratory performances (IN2 Group Ltd., Zagreb, Croatia). Medical laboratories receiving the questionnaire were instructed to ask their laboratory manager or laboratory professionals (or quality control manager) to fill it out.
The questionnaire was part of Module 11, entitled ‘Post-analytical phase of laboratory testing’, which explained post-analytical practices under the new Croatian EQA scheme. It consisted of closed-type questions covering four indicators of post-analytical quality proposed by the Working Group ‘Laboratory Errors and Patient Safety’ of the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) (5). The questions described specific situations or scenarios often encountered in routine practice concerning TAT, critical values, interpretative comments and procedures (repetition or additional testing) in the event of abnormal test results. Participants could choose only one of the offered responses for each question. Responses to 12 questions administered during one or more of the four exercises during the study period were analysed.
Results
The number of medical laboratories participating in each exercise varied, as did the number that filled out the questionnaire; Figure 1 presents a histogram of response rates to questions. The response rate was always more than 80% of laboratories participating in the exercise. Table 1 presents the contents of the questionnaire, which varied with the exercise.
Table 1
Responses to the questionnaire were analysed to determine to what extent medical laboratories comply with existing rules issued by professional bodies in Croatia, and to identify how medical laboratories are likely to proceed in frequent yet ‘grey’ situations where no clear rules exist. Frequencies of different responses to all questions are shown in Table 1 where answers that are clearly non-compliant with existing rules are emphasised in question comment. While laboratory respondents showed good knowledge of the definition of TAT, only approximately half of respondents reported that they routinely monitor TAT. Most respondents (85.7%) showed good knowledge of the definition of critical values (reflected in five questions) and of recommendations on how to apply the definition. On the other hand, the laboratories showed substantial variation in how they responded to certain questions about critical values; many lacked knowledge about age-dependent critical values and how to establish an intra-laboratory list of critical values.
The questionnaire presented respondents with three scenarios involving interpretative comments in order to understand to what extent medical laboratories in Croatia take an active role in interpreting test results and communicating those interpretations to clinicians orally or in writing. While laboratories varied in their responses to these scenarios, one third selected answers implying an active role in interpretation of test results, either via a comment written on the report or contact with the clinician and/or patient. In various situations, up to a third of laboratories issued results without additional activities.
The last group of questions asked laboratories how they proceed in the event of abnormal test results. For example, does the laboratory repeat the test and, if so, does it use the same or a new sample? Is the sample re-analysed using the same test procedure as before or a different procedure? After re-testing, are the initial and/or follow-up test results shown on the final report? Most laboratories reported that they repeat testing to verify abnormal results. Nevertheless, one third reported that they issue results without such verification.
Discussion
Harmonization and standardization of pre- and post-analytical phases of laboratory work are essential for good clinical care. Since 2007, ISO standard 15189 has included assessment of pre- and post-analytical phases of testing as one of the requirements for accreditation of medical laboratories (6). Nevertheless, many providers of EQA schemes do not systematically assess the post-analytical phase (2). Since 2014, all medical laboratories in Croatia are required to participate in a national EQA scheme that includes post-analytical assessment. The present study aimed to assess the current state of laboratory compliance with the EQA scheme, as well as identify areas where clearer rules - or the first set of rules - need to be developed at the national level. This is an urgent problem, because only 11 of 198 (5.5%) registered medical laboratories in Croatia are ISO 15189 - accredited, and most are planning to enter the accreditation process soon (7). In the present study, we retrospectively analysed the responses of medical laboratories to the Module 11 post-analysis questionnaire incorporated in the Croatian EQA exercises since 2014. This questionnaire focused on the four main quality control indicators of the post-analytical phase of testing. Our results indicate substantial heterogeneity in how medical laboratories in Croatia proceed in situations where no clear rules or guidelines exist.
TAT is a frequently used quality indicator: it is easily tracked through the laboratory informatics system, and ISO 15189 mandates that the TAT be established for each type of test, through consultation between laboratory and clinician (item 5.8.11) (6). One challenge with standardizing TATs across laboratories is that the definition of TAT can vary depending on whether the laboratory is a primary, secondary or tertiary facility and whether the test is routine, emergency or specialized (8, 9). Our results indicate that although most respondents know that TAT monitoring is an accreditation requirement, they do not have a monitoring system in place. Nevertheless, most respondents do record when the laboratory sample is received and when validated test results are obtained or reported. This likely reflects the widespread use of laboratory informatics systems.
ISO 15189 requires that laboratories apply standard procedures for recording and reporting critical values (items 5.8.1, 5.9.1 and 5.9.2) (6). Laboratories are also required to generate their own lists of critical values based on the local clinical situation, in consultation with clinicians (6, 10, 11). Most laboratories in our sample showed an understanding of critical values but not how to define own critical values list, or they neglected to adjust them based on patient age. In January 2015, CCMB published an updated and revised, ISO 15189 - compliant list of critical values (6, 12), which includes critical values and reference intervals for neonatal patients (12, 13). The recent release of this information at the national level may help to explain the heterogeneity in laboratory responses on our questionnaire. This information may help laboratories define their own lists of critical values and report critical values appropriately. Future work is needed to track laboratory - level implementation of this knowledge around the country.
Most medical laboratories reported that they confirm critical values with additional measurement before reporting, which is consistent with CCMB recommendations. Approximately one fifth indicated that they immediately report critical values without test repetition, which is consistent with practices at accredited laboratories in other countries (10, 14, 15) and reflects the fact that ISO 15189 - compliant laboratories may choose not to re-test in specific circumstances, such as when national guidelines in their country recommends it or when the use of advanced laboratory technology places the initial result beyond reasonable doubt. Recent published results of survey on critical results reporting in Croatian medical laboratories found high score for re-analyse critical results before reporting (16). In general, Croatian laboratories are in compliance with valid CCMB recommendation.
While those results based on a carefully designed scoring system are difficult to compare with our preliminary, descriptive results, the two studies may point to the need for more systematic research in this area.
Interpretative comments on the laboratory test report can improve treatment outcomes (17). They are a widely used quality indicator and, since 2007, an obligatory part of ISO 15189 accreditation (6). Although the CCMB has stated since 2004 that ‘remarks related to the sample (lipemia, hyperbilirubinemia, haemolysis and others) are a mandatory part of every report of laboratory test results’, the type, format and position of the comments on the report are not clearly defined (13), nor do CCMB guidelines indicate which comments necessitate contacting the clinician. Our results indicate that, depending on the situation, 3-35% medical laboratories do not flag abnormal results to the clinician, either in writing or orally. In addition, one third of laboratories neither repeats the test nor performs additional actions in an effort to confirm the abnormal results. Abnormal results may be significant for diagnosis and treatment, and may call into question the reliability of the test results. Including interpretative comments on lab reports can help prevent the release of incorrect or less reliable test reports (18-20). Therefore, our results identify an urgent need to revise and update CCMB recommendations about interpretative comments on test reports, as well as a need for the CCMB and other groups to define when tests or sampling should be repeated or additional tests performed.
A small proportion (12%) of respondent laboratories left open-ended comments to one or more of the questions; nearly all these comments were that their laboratory did not routinely encounter, or had never encountered, the scenario described in the question. This suggests that many laboratories feel they lack sufficient knowledge or experience to deal adequately with many post-analytical problems, despite the implementation of the Croatian EQA scheme. This suggests the need for greater training opportunities for medical laboratories in the country.
The present study presents a preliminary picture of the early stages of post-analytical EQA at the national level in Croatia. It is based on a sampling of medical laboratories from around the country and makes use of a questionnaire tailored to the logistical, clinical, and regulatory situation in Croatia. As with most questionnaire assessments, there is some risk that practices reported on the survey do not reflect actual practices in the respondent laboratory. To reduce this risk, we asked that the questionnaires be filled out by professional laboratory staff responsible for quality control. Another limitation of our study is that the response rate ranged from 81% to 90%, raising the possibility that our sample was biased. For example, perhaps laboratories that felt more confident about their knowledge and practices were more likely to respond to our survey. If this is true, then our study may underestimate the lack of alignment with post-analytical best practices, which only reinforces our conclusion that much more needs to be done to accelerate the harmonization of post-analytical procedures in Croatia. A third limitation is that the survey was not extensive enough to offer comprehensive insights into laboratory practices and attitudes. While this may have helped ensure a high response rate for the preliminary analysis here, future work may wish to look at these issues in greater detail.
In conclusion, assessment of post-analytical quality indicators such as TAT, critical values and interpretative comments are well recognized by both CCMB and ISO 15189, although clear definition of these terms, guidelines compliance and actions to be taken by laboratories are often incomprehensible. The results of Module 11 survey in Croatia highlights major obstacles to harmonization and standardization of post-analytical practices at national level. Future EQA exercises should reinforce the importance of filling out this survey.